Journal of Chemical and Pharmaceutical Research (ISSN : 0975-7384)

header
Reach Us reach to JOCPR whatsapp-JOCPR +44 1625708989
All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Original Articles: 2012 Vol: 4 Issue: 6

On-site determination of microalbuminuria based on ParticleEnhanced Turbidimetric-Inhibition Immunoassay (PETINIA) by portable fiber-optic spectrometer

Abstract

A homemade fiber-optic spectrometer has been fabricated for portable and on-site determination of microalbuminuria. The principle of the method was based on particle-enhanced turbidimetric-inhibition immunoassay (PETINIA), which was not affected by the hook effect phenomenon. This method was linear from 0 to 150 mg L-1 (r2 = 0.987), with a detection limit of 3.01 mg L-1 (S/N = 3). The intra-assay coefficient of variations (CVs) ranged from 3.1 - 4.8%, and the inter-assay CV was 6.7%, as assessed by calibration curves performed on three consecutive days. The precision of 10 replicate measurements of 50 mg L-1 HSA in a single analytical run was 6.5%. Interfering compound tests were satisfactory with recoveries of 95.0% to 105.8%. The system demonstrated its applicability to perform accurate measurements outside the laboratory for in-field analysis of the microalbuminuria (MAU) assay. The albumin concentrations in real urine samples measured by this system correlated well with the particle-enhanced turbidimetric immunoassay performed on a conventional spectrophotometer with r2 = 0.996 (n = 58).